Work Item
ASTM WK87954

Revision of F2477-23 Standard Test Methods for in vitro Pulsatile Durability Testing of Vascular Stents and Endovascular Prostheses

Rationale

FDA reviewers noticed an error in the radial force plots and a need for various clarifications. It was also noticed that equations for calculating diametric strain with constant length were missing.

Details

Developed by Subcommittee: F04.30

Committee: F04

Staff Manager: Kathleen Chalfin

Work Item Status

Date Initiated: 09-21-2023

Technical Contact: Brian Choules

Item: 000

Ballot:

Status: